SECTION 8—ACCESSORIES
Part No 1078988 33 Electric Portable Patient Lift - RPA6001E
Installing the Reliant Scale
NOTE:Forthisprocedure,refertoFIGURE 8.2.
WARNING
Patient and sling MUST be removed from the lift during ALL installation proce-
dures.
1. PositiontheloadcellassemblyoftheReliantScaleintotheboommountingbracket.
RefertoDetailʺAʺinFIGURE 8.2.
NOTE:Use1/4inchnylonwasher.
2. SecuretheReliantScaletotheboommountingbracketwithashoulderbolt,two��nylon
washersandalocknut.Securelytighten.RefertoDetailʺAʺinFIGURE8.2forwasher
orientation.
NOTE:Ensuretheshaftoftheshoulderboltpassesthroughbothsidesoftheboommounting
bracket.
3. Insertswivelbarpinwithtwonylonwashersthroughtheswivelbar.RefertoDetail
ʺBʺinFIGURE 8.2.
4. Alignthemountingholesintheswivelbarpinwiththemountingholesintheload
cellassembly.RefertoDetailʺBʺinFIGURE8.2.
5. Secureswivelbarpintotheloadcellassemblywiththeprovidedmountingscrewand
locknut.Securelytighten.RefertoDetailʺBʺinFIGURE 8.2.
WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may occur.
6. Weighthepatient.RefertoOperatingtheScaleonpage34andWeighingthePat ient
onpage 35.
FIGURE 8.2 Installing the Reliant Scale
Locknut
Swivel Bar
Nylon
Washer s
Swivel Bar Pin
Mounting
Screw
Boom
Reliant
Scale
Boom
Mounting
Bracket
Reliant
Scale
Nylon
Washer
Load Cell
Assembly
Shoulder
Bolt
DETAIL “B”
DETAIL “A”
Locknut
Nylon
Washer