SECTION 8—ACCESSORIES
Part No 1078988 37 Electric Portable Patient Lift - RPA6001E
3. PresstheCALbuttonlocatedonthePCboard.TheCALswitchisnotlabeledbutis
theonlybuttononthePCboardlocatedinthelowerrightcorner.OncetheCAL
buttonispushedʺCAL1ʺwillbeseeninthedisplaywindow.
NOTE:CAL1selectsthecalibrationmodeusing50poundsofcalibratedweight.CAL2selectsthe
calibrationmodeforusewith200poundsofcalibratedweight.PressingtheUNITSkeytoggles
betweenCAL1andCAL2.
4. Whenthedesiredcalibrationmodeisdisplayed,presstheLOCK/UNLOCKkey.
NOTE:ThedisplaywindowwillnowshowUnLd.
5. EnsurethatthereisnoloadonthescaleandpresstheZEROkey.
6. Thedisplaywindowwillshowadashedline(‐‐‐‐‐‐)scrollingacrossandthentheword
LOADwillappearinthedisplaywindow.
7. Performoneofthefollowing:
A. FORCAL1OPTION‐support50poundsofcalibratedweightfromthescaleand
presstheZEROkey.
B. FORCAL2OPTION‐support200poundsofcalibratedweightfromthescaleand
presstheZEROkey.
8. Thedisplaywindowwillshowadashedline(‐‐‐‐‐‐)scrollingacrossandthentheword
DONEwillappearinthedisplaywindow.
9. RemovetheweightfromthescaleandpresstheLOCK/UNLOCKkey.
NOTE:Theunitwillnowbeinatemporarytestmodeandwillnotlocktheweightdisplay.This
willallowweighttobeloadedandunloadedtocheckthecalibration.
10. TurntheunitoffbypressingtheON/OFFkey.
11. TurntheunitonbypressingtheON/OFFkey.
NOTE:Scalewillnowbeinnormaloperation.
Troubleshooting

Display Codes

CALIBRATIONREQUIRED‐Indicatesimproperstoredcalibrationdata,calibrationis
necessary.
OVERCAPACITY‐Indicatesaweightexceedingthecapacityhasbeenloadedonthescale.
SYMPTON PROBABLE CAUSE SOLUTIONS
Unit does NOT work properly. Battery failure. Check battery. Replace if necessary.
Battery has been replaced and unit
still does NOT work properly.
Contact Invacare for Service at 1-800-333-
6900

CAL

OCAP