Voltage requirements
AC power quality and conditioning
Environmental requirements
Dimensions and weight
∙Impedance between neutral and ground wires: less than 2 ohms
∙Neutral to ground voltage: less than 1 volt AC maximum
∙Voltage drop across circuit breaker: less than 0.1 volts
∙Impluse voltage spikes transients: less than 50% of nominal line voltage
∙Surge voltage: less than 5’1 of nominal line voltage
In some locations, even though recommended wiring practices have been followed, the power reaching the scanner may exhibit unstable power quality characteristics. System errors and intermittent data loss could result.
Power conditioning equipment may be required to overcome the power quality problems. Kodak can make arrangements for an
Consult with your electrical contractor or utility company to ensure that on- site wiring complies with the specifications and that wall outlets and comply with codes
This equipment can operate in a typical office environment under the following conditions:
Scanner | Temperature | Relative | Heat |
|
| Humidity | Dissipation |
9500/990 | 59° to 86° F | 15 to 76% | 5200 Btu |
| 15° to 30° C |
|
|
This equipment has the following dimensions and weight, with all optional accessories included:
Scanner | Height | Width | Depth | Weight |
9500 | 48 in. | 29 in. | 51 in. | 630 lbs |
| 122 cm | 74 cm | 130 cm | 227 kg |
990 | 48 in. | 29 in. | 51 in. | 785 lbs |
| 122 cm | 74 cm | 130 cm | 356 kg |
NOTE: With the Workstation Side Console, the width is 60 in (152 cm).
Space requirements Adequate work space must be provided to allow normal operation, proper