Scanner Time

Scanner Time

Actual (GMT): displays the Greenwich Mean Time of the scanner.

Corresponding local time: displays the scanner’s Greenwich Mean Time in the PC’s local time zone.

Refresh: displays the scanner time.

Calibrate — displays the Calibration window allowing you to perform an Image or UDDS calibration.

Image calibration: optimizes your scanner to achieve the best image quality and feeding performance. Frequent calibration is not necessary or recommended.

UDDS calibration: ensures that the ultrasonics system that detects multi-feeds and document edges is properly adjusted for best performance. Frequent calibration is not necessary or recommended.

Use the calibration target provided with your scanner. Be sure to use a good, clean calibration target.

1. Click Calibrate. The Calibration dialog box will be displayed.

2.Select Image or UDDS from the Type drop-down box.

If you select Image, use the calibration target provided with your scanner. Be sure to use a good, clean calibration target. Place the calibration target in the input elevator face up and click OK.

IMPORTANT:If the White Background Accessory is installed, it must be replaced with the black background stirps, and the scanner must be restarted prior to calibration.

If you select UDDS, place a 20 lb. or 75 g.m2 bond sheet of paper in the input elevator. Do not use the calibration target and click OK.

A-61504 April 2007

3-39