<Setup procedure>

This section describes the procedure for installing the RANGE7/5 properly, and completing the preparations for measurement. Make preparations according to the following procedures:

1)Installing the RANGE7/5.

2)Set the RANGE7/5 at a position relative to the proper target, and mount the lens suitable in question as the target of measure. (No replacement lens for RANGE5.)

3)Connect the AC adaptor to the RANGE7/5.

4)Connect the RANGE7/5 to a PC.

5)Turn ON the RANGE7/5 power switch to start warming up the RANGE7/5.

1) Installing the RANGE7/5

The RANGE7/5 can be directly installed on a table by using the rubber soles at the bottom. Also, it can be installed with the tripod set or measuring stand (under optional accessories). When the tripod set or measuring stand are used, the RANGE7/5 orientation can be freely changed (up/down and right/left). This feature is useful for measurement.

Memo For installation procedure details, please refer to the description on “Installation”.

2) Setting the RANGE7/5’s position relative to the measuring target, and exchanging the lens (RANGE7 only)

Scan Position Adjustment

The location and posture of the work and RANGE7/5 can be adjusted so that the two are appropriately positioned for scanning. These adjustments are made by activating the monitoring feature and watching the work in the monitor window.

Receiver lens RANGE7: TELE/WIDE replaceable RANGE5: No replacement lens

RANGE7/5

Distance to work

 

450mm to 800mm

X Z

Y

Work

The origin is set at the sensor position which is approximately 120 mm behind the center of the lens surface. Therefore, the Z dimension has an offset of around 120 mm applied.

RANGE VIEWER monitor window

Click the [Monitor] button to check the work on the monitor window.

Distance to work and work size

 

(Unit: mm)

 

 

 

 

 

Lens

TELE lens

WIDE lens

Distance

450

800

450

800

X×Y

79×99

141×176

150×188

267×334

Z

54

97

109

194

Memo For RANGE5, please refer to the WIDE lens data.

Distance to work and work size

 

 

in MultiFocus mode

 

 

(Unit: mm)

Lens

TELE lens

WIDE lens

Distance

462

781

475

766

X×Y

81×102

138×172

159×199

256×320

Z

54

97

109

194

Memo For RANGE5, please refer to the WIDE lens data.

12