<Setup procedure>
This section describes the procedure for installing the RANGE7/5 properly, and completing the preparations for measurement. Make preparations according to the following procedures:
1)Installing the RANGE7/5.
2)Set the RANGE7/5 at a position relative to the proper target, and mount the lens suitable in question as the target of measure. (No replacement lens for RANGE5.)
3)Connect the AC adaptor to the RANGE7/5.
4)Connect the RANGE7/5 to a PC.
5)Turn ON the RANGE7/5 power switch to start warming up the RANGE7/5.
1) Installing the RANGE7/5
The RANGE7/5 can be directly installed on a table by using the rubber soles at the bottom. Also, it can be installed with the tripod set or measuring stand (under optional accessories). When the tripod set or measuring stand are used, the RANGE7/5 orientation can be freely changed (up/down and right/left). This feature is useful for measurement.
Memo For installation procedure details, please refer to the description on “Installation”.
2) Setting the RANGE7/5’s position relative to the measuring target, and exchanging the lens (RANGE7 only)
Scan Position Adjustment
The location and posture of the work and RANGE7/5 can be adjusted so that the two are appropriately positioned for scanning. These adjustments are made by activating the monitoring feature and watching the work in the monitor window.
Receiver lens RANGE7: TELE/WIDE replaceable RANGE5: No replacement lens
RANGE7/5 | Distance to work |
| 450mm to 800mm |
X Z
Y
Work
•The origin is set at the sensor position which is approximately 120 mm behind the center of the lens surface. Therefore, the Z dimension has an offset of around 120 mm applied.
RANGE VIEWER monitor window
Click the [Monitor] button to check the work on the monitor window.
Distance to work and work size |
| (Unit: mm) | ||
|
|
|
|
|
Lens | TELE lens | WIDE lens | ||
Distance | 450 | 800 | 450 | 800 |
X×Y | 79×99 | 141×176 | 150×188 | 267×334 |
Z | 54 | 97 | 109 | 194 |
Memo For RANGE5, please refer to the WIDE lens data.
Distance to work and work size |
|
| ||
in MultiFocus mode |
|
| (Unit: mm) | |
Lens | TELE lens | WIDE lens | ||
Distance | 462 | 781 | 475 | 766 |
X×Y | 81×102 | 138×172 | 159×199 | 256×320 |
Z | 54 | 97 | 109 | 194 |
Memo For RANGE5, please refer to the WIDE lens data.
12