nAbout Field Calibration

The figure below shows an example of setting up the RANGE7/5 for calibration using the Calibration Set.

Enough space is needed to spread out the sheet.

 

 

1280×430 mm

 

 

 

Calibration chart

RANGE7/5

Set the calibration chart in the position

 

 

 

marked on the sheet for the lens being used.

 

 

Memo

 

Calibration base

 

 

For RANGE5, place in the

 

 

position off the WIDE lens.

 

 

Calibration sheet

 

 

 

Do not touch the front side of the chart with bare

 

hands. If dirty, wipe the chart clean with a dry

 

soft piece of cloth.

Calibration is easy to understand and perform

 

 

with the Calibration Wizard. For details, see “5-1.

Scanning is done seven times for calibration.

Calibration” on p. 70 of the Reference Manual.

Position and angle the calibration chart as

 

indicated in the Calibration Wizard.

Calibration should be performed in the following cases:

Before using for the first time after purchase.

If the RANGE7/5 has not been used for a few days.

If the ambient temperature during storage or use is drastically different from the ambient temperature at the time calibration was last performed.

If the RANGE7/5 may have been subject to vibrations during shipping or movement.

To achieve higher measurement accuracy.

NOTE

The Calibration Chart has been calibrated using the RANGE7/5 unit with which it was shipped. When performing calibration, be sure to use the Calibration Chart with the same number as the RANGE7/5 body.

Before performing calibration, confirm that the lens number is the same number as the RANGE7/5 body.

16