18
The following graticules and stage micrometers
are available for calibration:
Graticule 10mm/0.1mm
Graticule 5mm/0.1mm
Graticule 5mm/0.05mm
Graticule 100 div./0.002"
Graticule 100 div./0.001"
Graticule 150 div./0.0005"
Crosshair
Stage micrometer 50mm, 0.1/0.01mm divisions
Stage micrometer 1", 0.001" divisions
For photography, a 10×format reticle is available.
Two adjustable eyepieces are required for equipment
which has a reticle in an eyepiece for measurement or
photography purposes.
The graticules can be inserted into adjustable eyepieces
and eyepieces for spectacle wearers:
Ascertain with the aid of the stereomicroscope on
which side the vacuum-metalized scale is located.
The scale should be visible on the correct side.
Pull out the insert
from the bottom of
the eyepiece and
place it with the
knurled side on the
table.
Graticule
Adjustable
eyepieces
Assembly
Leica StereoZoom User Manual