29
Leica StereoZoom User Manual – Operating
Without looking into the eye-
pieces, turn the eyelens of the
adjustable eyepiece as far as
possible in the “+” direction.
Close the eye that is looking
into the fixed eyepiece and
look into the adjustable
eyepiece with the other eye.
View the test object and
slowly turn the eyelens clock-
wise (the “–” direction) until
the object is in focus.
Set the microscope to
maximum magnification.
View the test object with both
eyes and bring it into sharp
focus with the focusing drive.
View the object while zooming
from minimum to maximum magni-
fication.
The object should remain in
constant focus (parfocal) at all
times. If it does not, please
repeat the procedure above.
Diopter adjustment
at the adjustable
eyepiece
Checking
parfocality