CAUTION: If your Check Strip result falls outside the printed range, the Meter is not working properly. DO NOT use the Meter to test your blood until you get a Check Strip reading that is within the printed range. For assistance, call LifeScan Technical Services.
You must use the Check Strip:
■At least once a day.
■After cleaning the Meter.
■Whenever your results are not consistent with how you feel, or when you think your results are not accurate.
■Whenever this message appears:
A 2.1 (example)
Anotok redo A
Cleaning and Maintaining the Check Strip
■Make sure both sides of the Check Strip are clean. If not, wipe with a soft cloth or tissue dampened with water. Dry completely.
■Do not put blood, alcohol, Control Solution, or any other fluid (except water) on the Check Strip.
■Do not scratch the Check Strip.
■Do not leave the Check Strip in sunlight for long periods of time.