MESSAGE

PROBLEM

WHAT TO DO

not enough blood retest

not ok

A 5 .8 (example)

Anotok redo A

1)The blood sample was too small or smeared.

2)The Test Strip was not inserted all the way into the Test Strip Holder.

Your Meter may have an electronic problem.

The Check Strip test result is outside the acceptable range.

Repeat the test with a new Test Strip and a large, shiny drop of blood.

Repeat the test. This time push the Test Strip all the way into the Test Strip Holder.

Call LifeScan Technical Services.

Clean the Meter and Check Strip. Repeat the Check Strip test.

50

Page 57
Image 57
Lifescan blood glucose monitoring system manual Not enough blood retest Not ok, Anotok redo a