MESSAGE | PROBLEM | WHAT TO DO |
3)The Test Strip was inserted before insert strip appeared on the display.
4)The center of the Test Strip Holder where the small hole is located is raised.
Repeat the test. Wait for insert strip before inserting a new Test Strip.
Depress the center by pressing forward on the base, or by pressing down on the center, or by inserting a check strip into the Test Strip Holder.
c 3 .8 (example)
contrl
If this message appears after a blood test, your blood sample was too small, smeared, or another drop was added after the test began.
Repeat the test with a new Test Strip and a larger drop of blood.