Lincoln 1600 Temp. display, Display Transformer, inoperative, Temp. Display, Thermocouple

Models: 1600

1 28
Download 28 pages 62.83 Kb
Page 9
Image 9

Temp. display

Display Transformer

Measure the transformer primary for a nominal 120 VAC input,

inoperative

 

if voltage is not present, trace wiring back to the oven power

 

 

source. Measure the secondary output, which will normally be

 

 

12-15 VAC. Voltage across terminals 1 and 3 on the temp.

 

 

display should read 12-15 VAC. Terminals 1-2 and 2-3 should

 

 

be 1/2 the voltage reading to 1-3. If the above secondary

 

 

readings are not achieved, replace transformer.

 

Temp. Display

If proper voltage is present at terminals 1, 2, and 3 on

 

 

temperature display, but the display is not lighted, recheck all

 

 

connections for tightness. If temperature display is still not

 

 

operating, replace the temperature display.

Temp. Display

Thermocouple

Connect the thermocouple of a pyrometer to the thermocouple

inaccurate, erratic or

 

on the temperature control board. (Be sure to maintain proper

inoperative

 

polarity of thermocouple leads--see Adjustment Section for

 

 

proper procedure.) The pyrometer will indicate oven

 

 

temperature. Measure DC millivolts of thermocouple at the

 

 

temperature display. See thermocouple chart in adjustment

 

 

section for proper readings. Replace thermocouple if proper

 

 

readings are not achieved. If readings are correct on the

 

 

thermocouple, replace temperature display.

Low Profile – 1600-000-A Advantage Series Service Manual – Domestic

9

Page 9
Image 9
Lincoln 1600 Temp. display, Display Transformer, inoperative, readings are not achieved, replace transformer, Thermocouple