Table 2 Analog Outputs

Parameter

Oxygen

Possible

Possible

Name

Default

Options

Ranges

OUTPUT 1

O2

O2, CARBON,

O2 = 0 – 9999

MODE

 

DEWPT, TEMP, LIN,

%C = 0.00 – 2.55

 

0–20.9%

PROG

DP = -99.9 – 212.0

 

4-20mA

 

Temp = -999 – 3000

 

 

 

LIN = -999 – 9999

 

 

 

PROG = 0 – 4095

OUTPUT 2

TEMP

O2, CARBON,

O2 = 0 – 9999

MODE

 

DEWPT, TEMP, LIN,

%C = 0.00 – 2.55

 

800-3000°F

PROG

DP = -99.9 – 212.0

 

4-20mA

 

Temp = -999 – 3000

 

 

 

LIN = -999 – 9999

 

 

 

PROG = 0 – 4095

NOTE: SEE PAGE 4 FOR TYPICAL CALIBRATION VALUES.

Calibration

The Smart Transmitter is factory calibrated. The calibration can be verified once a year or according to customer calibration schedules. The instrument should be returned to the factory if calibration is required.