Thermocouple Null

 

Thermocouple Span

 

Cold Junction Trim

Communications port RS-485 Half Duplex Only

Protocol

Modbus RTU

Baud rates

1200, 2400, 4800, 9600, 19.2K (19.2K default)

Parity

None

Address

1 – 254 (Address 1 is default)

Housing

 

Material

Polyamide PA non-reinforced

Inflammability

Evaluation Class V0 (UL94)

Temperature Range

-40 to 100°C

Dielectric Strength

600 kV/cm (IEC243-1)

Mounting

Snaps on to EN 50022 top hat (T) style DIN rail.

Terminals

Wire clamp screw terminals on four position removable terminal blocks.

Wire Size

AWG 24 – 12 flexible stranded, removable terminal blocks.

Max. Torque

0.8 Nm

CAUTION: DO NOT CONNECT OR DISCONNECT HOUSING PLUGS WHILE MODULE IS POWERED OR UNDER LOAD.

Weight

10 oz

 

 

Environmental Conditions

 

 

Operating Temperature

-20

°C to 55 °C (-4 to 130 F)

Storage Temperature

-40

°C to 85 °C (-40 to 185 F)

Operating and Storage Humidity

 

 

 

85% max relative humidity, noncondensing, from –20

 

 

to 65°C

Certifications and Compliance (PENDING)

Safety

EN 61010-1, IEC 1010-1

Safety requirement for electrical equipment for measurement, control, and laboratory use, Part 1

Electromagnetic Compatibility

Immunity as specified by EN 50082-2

Electrostatic discharge

EN 61000-4-2

Level 3: 8 kV air

Electromagnetic RF fields

EN 61000-403

Level 3: 10 V/m

 

 

80 MHz – 1 GHz