Program Description
Custom Openprogram available for programming by end user.
INORGANICA Forremoving inorganic residue;
generallyfor analytical applications, for light to moderate soiling, with moderate
tohigh rinse requirements.
ORGANICA Forremoving organic residue, such as oils, fats, wax, agar, preparation
applicationsand for some analytical applications, with moderate to heavy soiling,
andmoderate rinse requirements. Liquid detergent required, cold, warm, and a
DIwater supply is recommended.
STANDARD Fora wide range of soiling types that do not denature or precipitate in alkaline
environments,for some preparation and analytical applications, for light soiling,
andlow rinse requirements.
UNIVERSAL Forremoving organic residue (such as proteins, oils and some fats) and some
inorganicresidues (pH 7, water-soluble metallic salts), for preparation
applicationsand for some analytical applications, moderate to heavy soiling and
moderaterinse requirements.
INTENSIVE Forremoving organic residue (such as proteins, cell and tissue cultures, oils and
somefats) and some inorganic residues (pH 7, water-soluble metallic salts), for
preparationapplications and analytical applications, moderate to heavy soiling,
andmoderate to high rinse requirements.
PLASTIC Fortemperature-sensitive labware, such as plastic bottles (minimum required
temperatureresistance 55°C, for preparation applications and for some
analyticalapplications, for light to moderate soiling, for moderate rinse
requirements.
DRAIN Topump away water, for example when a program is canceled.
RINSE Torinse heavily soiled labware, e.g., to pre-treat heavy soiling, remove residue or
toavoid drying and encrusting before a full program is run.
varioTD Forcleaning and thermal disinfection at 93 °C with 5 min holding time (as per
ENISO 15883-1).
SPECIAL
93°C - 10'
Forcleaning and thermal disinfection at 93 °C with 10 min holding time (as per
¶18 of Germany's infectious disease control act, section A/B, fungicidal,
bactericidal,and virus-inactivating).
CW= cold water, HW = hot water, DI = distilled water, °F (C°) = Temperature, min = holding time
*The DOS 2 rinse aid dosage must be activated by Miele Technical Service
Program Guide
32