Appendix E Common Questions
NI 2501/2503 User Manual E-2
©
National Instruments Corporation
Finally, verify that your code is correct. For reference, see the
examples described in the NI-SWITCH Software User Manual. The
NI-SWITCH driver also ships with several examples in source code.
Compare your algorithm to the ones in the examples.
What should I do if scanning does not work?
First, ensure that you have configured the switch card and the
instrument to match trigger lines. The output trigger of the instrument
should connect to the trigger input of the switch card. In addition, the
scanner advanced trigger of the switch card should be connected to the
input trigger of the instrument.
If the switch card is used to initiate the scan, make sure the DMM is
waiting for a trigger before enabling scanning on the switch card. This
is the recommended method for hardware scanning.
If the DMM is used to initiate the scan, enable scanning on the switch
card before configuring the DMM to start taking measurements.
Also be sure to check the return codes of the NI-SWITCH operation to
ensure that there are no errors (negative value) or warnings (positive
value). You can also use the NI-Spy utility to check for error codes.
Finally, verify that your code is correct. For reference, see the
examples described in the NI-SWITCH Software User Manual. The
NI-SWITCH driver also ships with several examples in source code.
Compare your algorithm to the ones in the examples.
Do I need to program the switch card myself?
The NI25 01/2503 comes with the NI-SWITCH driver software, which
exports the full functionality of the card. NI-SWITCH handles the complex
issues of direct memory access, interrupts, and operating system
interfacing.