© National Instruments Corporation 15 NI 5412 Specifications
MarkersTable 8.
Specification Valu e Comments
Destinations 1. PFI<0..1> (SMB front panel connectors)
2. NI PXI-5412—PXI_Trig<0..6> (backplane connector)
NI PCI-5412—RTSI<0..6>
—
Quantity One Marker per Segment. —
Quantum Marker position must be placed at an integer multiple of
four samples.
—
Width >150 ns. Refer to tm2 at NI Signal Generators Help»
Devices»NI 5412»NI <bus>-5412»Waveform
Generation»Marker Events.
—
Skew
Destination
With Respect to Analog
Output
Refer to tm1 at
NI Signal
Generators
Help»Devices»
NI 5412»
NI <bus>-5412»
Waveform
Generation»
Marker Events.
PFI<0..1> ±2 Sample Clock Periods
NI PXI-5412
PXI_Trig<0..6>
NI PCI-5412 RTSI<0..6>
±2 Sample Clock Periods