Specifications

 

Item

Specifications

Laser light

Visible light laser source

Select from the following laser, to mounted on laser combiner

source

 

Multi-line Ar laser (458nm, 488nm, 515nm, Total 40mW), Ar laser (488nm,10mW), Kr laser (568nm, 10mW),

 

 

HeNe (G) laser (543nm,1mW), HeNe (R) laser (633nm,10mW), LD405 (405nm, 25mW), LD440 (440nm, 5.3mW)

 

 

 

 

Laser combiner

Each laser light path is equipped with a continuously variable neutral density filter or AOTF

 

 

All laser lines are combined to apsis along the same fiber optic

 

 

 

Scanning unit

Scanning method

Galvanometer mirror scanners (both X and Y)

 

 

 

 

Field number

20 (10 with use of LD405 laser)

 

 

 

 

Pinhole

5-position pinhole turret

 

 

 

 

Image memory and

Standard scanning mode: 256 x 256 (0.45s) - 2048 x 2048 (10.835s)

 

scanning speed

Bi-directional high-speed scanning mode: 512 x 512 (0.25s) (Simultaneous scanning of up to 2 channels)

 

 

 

 

Image channel

Selectable from 2-channel (fluorescence) or 2-channel (fluorescence) + 1-channel (transmitted light)

 

 

3-channel (fluorescence) using virtual channel

 

 

 

 

Selection of filters according to staining

Manual selection

 

 

 

 

Scanning modes

XY, XYZ, XYT, XYZT, XZ, XT, XZT, point, Line-t, free line-t, line-z, free line-z, Clip, ZoomIn

 

 

 

 

Image depth resolution

12-bit (=4096 grey levels)

 

 

 

 

Zoom

1X-10X

 

 

 

 

Z-drive

Step motor/Minimum step 10nm (BX61, BX61WI and IX81 combination), 25nm (other microscope combination)

 

 

 

Microscopes

Upright

BX51, BX61, BX51WI, BX61WI

 

 

 

 

Inverted (special laser safe frame)

IX81FVSF, IX71FVSF (side port)

 

 

 

External trans-

Transmitted light illumination unit

External halogen light source connected to microscope via fiber cable

 

 

 

mitted light unit

Transmitted light detector

External detector unit with built-in photomultiplier Connected to microscope frame via fiber cable

 

 

 

Fluorescence illumination unit

Connect to external mercury light source and microscope via fiber cable

 

 

Standard equipment of FV300-BX51, FV-300-BX61, FV300-BX51WI, FV300-BX61WI

 

 

 

PC with system control boards

PC-AT compatible machine/OS: Windows XP (English version)/ 1GB memory (can be expanded to a maximum of 4GB)

 

 

CPU: Pentium 4, over 2.8GHz, Special I/F board/image capture: PCI bus

 

 

Graphic board: G450_Dual 32MB

 

 

Hard disk: 80GB 7200rpm_ID (ATA100) with DVD-ROM

 

 

Monitor: Two 19” LCD monitors are recommended, each able to display 1280x1024 images in full color (16.77 million colors)

 

 

LAN: On board

 

 

 

Fluoview

Image acquisition

Scanning condition setting: image size, scanning speed, zoom, panning etc.

application

 

Real-time image calculation: Kalman filtering, peak integration,

software

 

 

Hardware control

Laser, scanning unit, microscope

 

 

 

 

 

 

Each image display: Single-channel side-by- side, merge, cropping, tiling, series (Z/T) pass and continuous

 

Image display

LUT: Individual color setting, pseudo-color, Overlay: Lines, text, scale bar, etc

 

 

 

 

Image processing

Individual filter: Average, Low-pass, High-pass, Sobel, Median, Prewitt, 2D Laplacian, edge enhancement etc.

 

 

Calculations: Inter-image, mathematical and logical,

 

 

DIC back ground leveling

 

 

 

 

Image analysis

Overview of fluorescence intensity within an area, histogram, perimeter measurement for user-assigned area,

 

 

time-lapse measurement , etc.

 

 

 

 

3D visualization

3D animation, left / right stereo pairs, red / green stereoscopic images and cross section

 

 

 

 

Others

Graphic-based help, PAPP (Programmable Acquisition Protocol Processor), time course software (optional),

 

 

trigger IN/OUT function (optional), Multi point software (optional)

 

 

 

Power consumption

Microscope (115V 6A/230V 3A), scanning unit+PSU (115V 3.5A/230V 2A), computer & monitor (115V 4.5A/230V 10A),

 

 

Ar laser (115V 10A/230V 5A), Mult-linei Ar laser (115V 10A/230V 5A), Kr laser (230V 20A),

 

 

HeNe laser each (115V 0.4A/230V 0.2A), LD laser (405nm, 440nm: 100V 0.9A/230V 0.5A)

 

 

 

Objectives for BX and IX

(using U-UCD8, IX-LWUCDA and U-DICTS)

 

 

 

Cover glass

 

 

Condenser for BX

Condenser for IX

U-DICTS

Description

NA

W.D

Immersion

Correction ring

U-UCD8A

IX-LWUCDA

thickness

position

 

 

 

 

 

optical element

optical element

 

 

 

 

 

 

 

UPLSAPO 4X

0.16

13

 

 

 

 

 

 

 

 

 

 

 

 

 

 

UPLSAPO 10X

0.40

3.1

0.17

 

 

U-DIC10

IX2-DIC10

normal

 

 

 

 

 

 

 

 

 

UPLAPO 10XO3

0.40

0.24

0.17

Oil

 

U-DIC10

IX2-DIC10

normal

 

 

 

 

 

 

 

 

 

UPLAPO 10XW3

0.40

0.43

0.17

Water

 

U-DIC10

IX2-DIC10

normal

 

 

 

 

 

 

 

 

 

UPLSAPO 20X

0.75

0.6

0.17

 

 

U-DIC20

IX2-DIC20

normal

 

 

 

 

 

 

 

 

 

UPLAPO 20XO3

0.80

0.19

Oil

 

U-DIC20

IX2-DIC20

normal

 

 

 

 

 

 

 

 

 

UPLSAPO 40X

0.90

0.2

0.11-0.23

 

_

U-DIC40

IX2-DIC40

normal

 

 

 

 

 

 

 

 

 

UPLFLN 40XO

1.30

0.2

0.17

Oil

 

U-DIC40

IX2-DIC40

normal

 

 

 

 

 

 

 

 

 

PLAPON 60XO

1.42

0.15

0.17

Oil

 

U-DIC60

IX2-DIC60

BFP1

 

 

 

 

 

 

 

 

 

UPLSAPO 60XO

1.35

0.15

0.17

Oil

 

U-DIC60

X2-DIC60

normal

 

 

 

 

 

 

 

 

 

UPLSAPO 60XW

1.20

0.28

0.15-0.2

Water

_

U-DIC60

X2-DIC60

normal

 

 

 

 

 

 

 

 

 

UPLSAPO 100XO

1.40

0.12

0.17

Oil

 

U-DIC100

X2-DIC100

normal

 

 

 

 

 

 

 

 

 

Objectives for fixed stage upright microscopes (using WI-UCD, WI-DICTHRA)

Objective

N.A.

W.D.

DIC

Revolving

 

 

 

prism

nosepiece

 

 

 

 

 

MPL5X

0.10

19.60

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

UMPLFL10XW

0.30

3.30

U-LDPW10H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

UMPLFL20XW

0.50

3.30

U-LDPW20H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

LUMPLFL40XW

0.80

3.30

U-LDPW40H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

LUMPLFL60XW

0.90

2.00

U-LDPW60H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

LUMPLFL40XW/IR

0.80

3.30

U-LDPW40H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

LUMPLFL60XW/IR

0.90

2.00

U-LDPW60H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

LUMPLFL100XW

1.00

1.50

U-LDPW60H

WI-SSNP,

 

 

 

 

WI-SRE2

 

 

 

 

 

XLUMPLFL20XW

0.95*

2.00

U-LDPXLU20

WI-SNPXLU

 

 

 

HR

 

 

 

 

 

 

*Note: These conditions are not met in confocal microscopy

14

Page 15
Image 15
Olympus FV300 manual Specifications