Table 1: Meter-Mode Zero & Gain Adjustments

 

 

Sensor

 

 

U.U.T.

 

 

 

 

U.U.T.

Step

 

Select

 

 

Input

 

Adjust

 

Reading*

 

 

 

 

 

 

 

 

 

 

1

 

K

 

 

0.000 mV

 

P1

 

32.0 ±0.1°F

2

 

K

 

 

54.856 mV

 

P11

 

2500.0

±0.1°F

 

 

 

 

 

 

 

 

3

 

 

Repeat Steps 1 & 2 As Required

 

 

 

 

 

 

 

 

 

 

 

4

 

J

 

 

0.000 mV

 

P1

 

32.0 ±0.1°F

5

 

T

 

 

0.000 mV

 

P1

 

32.0 ±0.1°F

6

 

E

 

 

0.000 mV

 

P1

 

32.0 ±0.1°F

7

 

N

 

 

0.000 mV

 

P1

 

32.0 ±0.1°F

8

 

B

 

 

1.439 mV

 

P1

 

1000.0

±0.5°F

9

 

R

 

 

0.000 mV

 

P1

 

32.0 ±0.5°F

10

 

S

 

 

0.000 mV

 

P1

 

32.0 ±0.5°F

11

 

**G**

 

 

2.238 mV

 

P1

 

600.0 ±0.2 °F

12

 

C

 

 

0.000 mV

 

P1

 

32.0 ±0.2°F

13

 

D

 

 

0.000 mV

 

P1

 

32.0 ±0.2°F

 

 

 

 

 

 

 

 

 

 

 

** exclusive of noise.

 

 

 

 

 

 

 

 

 

** on EC models, G replaced by L. Calibrate L per J.

 

 

 

Table 2: Calibrator-Mode Zero & Gain Adjustments

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor

 

 

U.U.T.

 

 

 

 

DMM

 

Step

 

Select

 

 

Output

Adjust

 

 

Reading*

1

 

K

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

2

 

K

 

 

2500.0°F

 

P41

 

54.856 mV ±1µV

 

 

 

 

 

 

 

 

 

 

 

3

 

 

Repeat Steps 1 & 2 As Required

 

 

 

 

 

 

 

 

 

 

 

 

4

 

J

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

5

 

T

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

6

 

E

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

7

 

N

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

8

 

B

 

 

1000.0°F

 

P1

 

1.439 mV ±1µV

9

 

R

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

10

 

S

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

11

 

**G**

 

 

600.0°F

 

P1

 

2.238 mV ±1µV

12

 

C

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

13

 

D

 

 

32.0°F

 

P1

 

0.000 mV ±1µV

 

 

 

 

 

 

 

 

 

 

 

 

** exclusive of noise.

** on EC models, G replaced by L. Calibrate L per J.

26

Page 28