Omega manual CDTX-90-2 Calibrate Menu

Models: CDTX-90

1 20
Download 20 pages 4.31 Kb
Page 12
Image 12

CDTX-90-2 Calibrate Menu

Display

 

Description

(Factory settings shown)

 

 

 

 

Cell Constant:

 

Select CUSTOM only if you are connecting a certified conductivity sensor. Select

Standard

>

STANDARD for all other systems.

 

 

 

 

Cell: Standard

1 >

For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.

Cell: Custom

1.0000 >

For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor.

Cond Units:

S >

Select from these options: S, mS, k, M, PPM

PPM Factor:

2.00>

If the Units selection is PPM, set the ratio of Total Dissolved Solids to S. The maximum allowable is 3 S = 1 PPM. The TDS factor can only be set in PPM.

Set:

Temperature >

Adjust the temperature of the system based on an accurate external reference.

Set:

 

When using a h STANDARD sensor perform this single-point wet calibration for best

Conductivity

>

accuracy. Not required for CUSTOM sensors except for periodic calibration checks.

 

 

 

 

 

 

 

Loop Source:

 

Select the input (Conductivity or Temperature) source to be associated with the current

Cond

>

loop output:

 

 

 

 

Loop Range: S

 

Select the minimum and maximum values for the current loop output.

0.0000 100.000

>

Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

 

 

 

Relay1 Source:

 

Select the source (Conductivity or Temperature) for this relay output:

Cond

>

 

 

 

 

 

 

 

Relay1 Mode:

 

Select the mode of operation for the relay.

Low

>

Options available are High, Low or proportional Pulse.

The relay may be disabled (Off) if not in use.

 

 

 

 

 

 

Relay1 Setpnt:

 

In Low or High Mode, the relay will be activated when the process reaches this value. Be

10.0000 S

>

sure to modify this setting if you change the Cond. Units.

 

 

 

 

 

 

 

Relay1 Hys: S

 

The relay will be deactivated at Setpoint ± Hysteresis, depending on High or Low Setpoint

0.5000

>

selection.

 

 

 

 

 

 

 

Relay1 Rng: S

 

In Pulse mode, set the process values where the proportional pulse will start and where it

10.0000 40.0000

>

will reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

 

 

 

Relay1 PlsRate:

 

In Pulse mode, set the maximum rate for the proportional Pulse. The transmitter will

120 Pulses/Min

>

accept any value from 0 to 400.

 

 

 

 

 

 

 

Last CAL:

 

Use this “note pad” to record important dates, such as annual recertification or scheduled

06-30-01

>

maintenance.

 

 

 

 

page 12

Page 12
Image 12
Omega manual CDTX-90-2 Calibrate Menu