Omega manual CDTX-90-1 Calibrate Menu

Models: CDTX-90

1 20
Download 20 pages 4.31 Kb
Page 10
Image 10

CDTX-90-1 Calibrate Menu

Display

 

Description

(Factory settings shown)

 

 

 

 

Cell Constant:

 

Select CUSTOM only if you are connecting a certified conductivity sensor. Select

Standard

>

STANDARD for all other systems.

 

Cell: Standard

1 >

For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.

Cell: Custom

1.0000 >

For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor.

Cond Units:

uS >

Select from the following options: S, mS, k, M, PPM

PPM Factor:

2.00>

If the Units selection is PPM, you can set the ratio of Total Dissolved Solids to S. See page 15 for additional information. The TDS factor can only be set in PPM.

Set:

Temperature >

Adjust the temperature of the system based on an accurate external reference.

Set:

 

For use with STANDARD sensors: Perform this single-point wet calibration for most accurate

Conductivity

>

results. Not required for CUSTOM sensors except for periodic system accuracy confirmation.

Enter all zeroes to restore factory calibration to TEMP and COND settings.

 

 

 

Loop Source:

 

Select the input source to be associated with the current loop output: Conductivity or

Cond

>

Temperature.

 

 

 

 

 

 

 

Loop Range: uS

 

Select the minimum and maximum values for the current loop output.

0.0000 100.000

>

Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

Output Source:

 

Select the source for the Open Collector output: Conductivity or Temperature

Cond

>

 

 

 

 

Output Mode:

 

Select the mode of operation for the Open Collector output.

Low

>

Options available are High, Low or proportional Pulse.

The signal may be disabled (Off) if not in use.

 

 

 

Output Setpnt:

 

In Low or High Mode, this Open Collector output will be activated when the process reaches

10.0000 uS

>

this value. Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

 

 

 

Output Hys:

 

The Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or

0.5000 uS

>

Low Setpoint selection. (See details in section 3.3.)

 

 

 

 

Output Rng: uS

 

In Pulse mode, set the process values where the proportional pulse will start and where it will

10.0000 40.0000

>

reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

Output PlsRate:

 

120 Pulses/Min

>

In Pulse mode, set the maximum rate for the proportional Pulse output. The CDTX-90 will accept any value from 0 to 400.

Last CAL:

06-30-01 >

Use this “note pad” to record important dates, such as annual recertification or scheduled maintenance.

page 10

Page 10
Image 10
Omega manual CDTX-90-1 Calibrate Menu