CDTX-90-3 Calibrate Menu

Display

 

Description

(Factory settings shown)

 

 

 

 

Cell Constant:

 

Select CUSTOM only if you are connecting a certified conductivity sensor. Select

Standard

>

STANDARD for all other systems.

 

Cell: Standard

1>

For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.

Cell: Custom

1.0000 >

For CUSTOM sensors: Enter the precise cell constant from the certificate provided with your sensor, or from the information label on the sensor.

Cond Units:

uS >

Select from the following options: S, mS, k, M, PPM

PPM Factor:

2.00>

If the Units selection is PPM, you can set the ratio of Total Dissolved Solids to S. See section 3 of this manual for additional information. The TDS factor can only be set in PPM.

Set:

Temperature >

Adjust the temperature of the system based on an accurate external reference.

Set:

 

For use with STANDARD sensors: Perform this single-point wet calibration for most accurate

Conductivity

>

results. Not required for CUSTOM sensors except for periodic system accuracy confirmation.

Enter all zeroes to restore factory calibration to TEMP and COND settings.

 

 

 

Loop1 Source:

 

Select the input source to be associated with the current loop output: Conductivity or

Cond

>

Temperature

 

 

 

 

 

 

 

Loop1 Range: uS

 

Select the minimum and maximum values for the current loop output.

0.0000 100.000

>

Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

Output Source:

 

Select the source for the Open Collector output: Conductivity or Temperature

Cond

>

 

 

 

 

Output1 Mode:

 

Select the mode of operation for the Open Collector output.

Low

>

Options available are High, Low or proportional Pulse.

The signal may be disabled (Off) if not in use.

 

 

 

 

 

 

Output1 Setpnt:

 

In Low or High Mode, this Open Collector output will be activated when the process reaches

10.0000 uS

>

this value. Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

 

 

 

Output1 Hys:

 

The Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or

0.5000 uS

>

Low Setpoint selection. (See details on page 4.)

 

 

 

 

Output1 Rng: uS

 

In Pulse mode, set the process values where the proportional pulse will start and where it will

10.0000 40.0000

>

reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.

 

 

 

 

Output1 PlsRate:

 

120 Pulses/Min

>

In Pulse mode, set the maximum rate for the proportional Pulse output. The transmitter will accept any value from 0 to 400.

Last CAL:

06-30-01 >

Use this “note pad” to record important dates, such as annual recertification or scheduled maintenance.

Settings will repeat for Loop 2 and Output 2.

page 14

Page 14
Image 14
Omega manual CDTX-90-3 Calibrate Menu