FOB101 | FOB102 or 104 | Description |
|
|
|
b | b | Return enclosure temperature |
N/A | e[1;2; ... ;n] | Enable channels to scan (if no argument, scan all) |
f[j] | f[i;j] | Set point adjustment on channel [i] to [j] |
h | h | Help menu (this screen) |
i | i | Get factory and status information |
l[i] | l[i] | Lock the keypad (l+ to lock, l- to unlock) |
o[i] | o[i] | Analog out, if error |
p[i] | p[i] | Sleep mode (p+ to sleep, p- for continuous scanning) |
q | q | Scanning rate (q+ for fast/quick, q- for slow) |
s[j] | s[i;j] | Specify analog output Span [j] for channel [i] |
t | t[i] | Get Temperature reading channel [i] |
ta[i] | ta[i] | Auto temperature output to serial port (+ to enable, - to disable) |
u[i] | u[i] | Unit (c = oCelcius, f = oFahrenheit) |
y | y | Quality index |
z[j] | z[i;j] | Temperature [j] for analog output Zero on channel [i]. |
|
|
|
5.3 Detailed description of RS-232 commands
All commands must be terminated by a carriage return character ( E ). When more than one argument
are required, each argument must be separated from the next one by a blank or space character. Characters are not echoed back to your controlling device.
Every time a command is executed successfully by the thermometer, the “*” prompt character is automatically sent, except in case of failures when the “Errx” error code is sent instead. Do not send your next command before receiving either the “*” or the “Errx” prompt. The “[“ and “]” characters must not be typed in; they are used here for ease of reading only.
Most commands are applicable to all FOB100 models (different number of channels). When using one- channel model, the channel number [i] must not be typed in.
“b” This command returns the internal temperature of the enclosure. The temperature is returned in oC or oF unit, as specified by the “u” command.
ex: | “b:32.2 E“ |
are enabled and read by your thermometer. Preceding a channel number with the minus sign
ex: |