To change the command set:

1.Press and hold the GPIB ADDR key while the SB switch powers up. The active interface setting is displayed.

2.Press the ▲ key to scroll to the required command set, for example, SB, SC, or SX.

3.Press the GPIB ADDR key to save the setting.

SC Series Switch Command Set (SCXXX-0XXX Models)

The syntax and functions for the SC series switch commands are provided in Table 12.

Table 12: SC Switch Command Set

Syntax

Function

 

 

AnE

switch to channel n

 

 

FE

verify channel

 

 

XE

turn external driver on

 

 

YE

turn external driver off

 

 

Response messages (Table 13) are formatted as qn, where q is the switch status (A = normal, I = syntax error, C = calibration error), and n is the channel number.

Table 13: Response Messages for SC Switch Commands

Status Byte Register

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Message

 

 

 

 

 

 

 

 

0

0

0

0

0

0

0

BUSY

 

 

 

 

 

 

 

 

0

0

0

0

0

0

1

NORMAL

 

 

 

 

 

 

 

 

0

RQS

0

0

0

0

1

CAL ERROR

 

 

 

 

 

 

 

 

0

RQS

0

0

1

0

1

BAD DATA

 

 

 

 

 

 

 

 

0

RQS

0

0

1

1

1

UNIT FAILURE

 

 

 

 

 

 

 

 