Omega WMS-16 manual Barometric Pressure Installation, Initialization Of The Barometer

Models: WMS-16

1 23
Download 23 pages 54.49 Kb
Page 16
Image 16

From the Station Set Up menu press number 6. ADD OR REMOVE SENSORS. Then press Y, then ENTER.

From the Sensor Type Menu select number 4. Temperature. Then press

ENTER.

From the Temperature Sensors Menu select number 1. (WMS-16T). Then press ENTER.

The menu then asks: Enter temperature offset or press Q to quit. Press Q.

From the Sensor Input Channel Menu select number 9. A5. Then press ENTER.

If you intend to add another sensor select A.

If you do not want to add another sensor select Q.

You will be asked if you are sure that you want to erase memory. If you have added or deleted a sensor select Y if not select N and press ENTER.

If you selected AA@ to add the Relative Humidity Sensor then press Y, then ENTER.

From the Sensor Types menu select 5. (Humidity)

From the Humidity Sensors menu select 1. (25500504(U)/40). Then press ENTER

From the Sensor Input Channel select 8. (A4). Then press ENTER

The Sensor Configuration menu appears and it will show:

 

Sensor type

Model

Input

Calibration

 

Humidity

WMS-16-PN

A4

 

Temperature

WMS-16T

A5

 

If you intend to add another sensor select A. If you do not want to add another sensor select Q.

You will be asked if you are sure that you want to erase memory. If you have added or deleted a sensor select Y if not select N and press ENTER.

Barometric Pressure Installation

Connect the barometer cable to the terminal strip on the back of the control module. This sensor has a three conductor cable. The RED wire is connected to a +12V terminal. The BLACK wire is connected to a GROUND terminal and the GREEN wire is connected to terminal A2.

INITIALIZATION OF THE BAROMETER

Press EXC to get the MAIN MENU.

From the MAIN MENU press number 1. STATION SETUP.

15

Page 16
Image 16