3Route. the sensor cable through a pipe of an appropriate material and length. Screw the compression fitting onto the end of the pipe (or coupling, if used).

Recommendation: Use Teflon tape or ipe sealant with Teflon.

(Exception: For higher temperature soPutions, sealing with Teflon tape may not be adequate.)

4.Fasten a Unilet junction box onto the other end of the pipe.

5.Run interconnect cable from the analyzer into the Unilet junction box. Connect the sensor and interconnect cable wires, by matching colors, to the terminals in the junction box. Fasten the cover onto the junction box.

6.Electrically connect the sensor interconnect cable wires to the analyzer. Refer

to the analyzer instruction manual for details.

7.Calibrate the analyzer using the procedure in the analyzer instruction manual.

8.After calibration, mount the sensor into the process.

This completes the immersion mounting.