Sensor Reading After Callouts

Sensor Reading After Callouts

If the Four Zone Temperature Monitor has called all programmed telephone numbers and not made contact it will wait 20 minutes before attempting to callout again. Periodically, the time before callouts resume will be displayed.

 

1: 510 >

CALL-

 

 

 

 

2: 200

OUTS

 

 

3:

IN

 

 

4:

17:25

 

 

 

 

 

 

 

 

 

18

Programming Sensor Parameters

1. Accessing the Sensor Configuration

a)From the Main Menu, press 2 to Set Limits

► The "Full Access" PIN will be requested if the "Acknowledge Only" PIN was entered initially. If the correct "Full Access" PIN is not entered, the Four Zone Temperature Monitor and Alarm will hang up.

b)You will hear “Enter Sensor Number"

c)Enter the sensor you want to program (1-4) ►To return to the Main Menu press 0

d)Proceed to Step 2.a.

2. Programming the Sensor ID message

a)You will hear "Sensor x message is …."

b)You will hear “Press 1 to change

c)Press 1 to change the message

► Press 2 to skip and proceed to step 3.a or press 0 to stop program- ming this sensor and return to step 1.b

d)You will hear a tone

1: 510 > Record something specific that will allow the people receiving the TIP alarm calls to understand where the problem is.

e)Begin speaking after the tone. The Four Zone Temperature Monitor and Alarm will record for about 3 1/2 seconds

f)After 3 1/2 seconds you will hear the tone again, marking the end of your message

g)You will hear the message you recorded

h)Proceed to step 3.a

3. Programming the Lower and Upper Temperature Limits

a)You will hear "Sensor x lower limit is" and the current low temperature limit for that sensor (i.e. 35º)

b)You will hear “Press 1 to change

c)Press 1 to change the limit

Press 2 to skip and proceed to step 3.g or press 0 to stop program- ming this sensor and return to step 1.b

d)You will hear “Enter number then press pound”

e)Enter the value then press #.

Use * to program a negative number (i.e. *20 = -20º)

Acceptable range is -999 to 999

f)You will hear the value you just entered (i.e. 39º)

g)You will hear "Sensor x upper limit is " and the current high temperature limit for the selected sensor (i.e. 60º)

h)You will hear "Press 1 to change"

i)Press 1 to change the limit

Press 2 to skip and proceed to step 3.g or press 0 to stop program-

7

Page 7
Image 7
Omega Engineering OMA-VM520-DCP Outs, Sensor Reading After Callouts, Programming Sensor Parameters