2Using the Handheld Infrared Thermometer

1.Mount the thermometer on a camera tripod and aim at the target.

2.Connect the analog output of the thermometer to a strip chart recorder as shown in Figure 2-11.

3.Pull the trigger and press the LOCK key to lock the trigger.

4.If necessary, adjust the emissivity.

5.The thermometer is now set up for unattended monitoring of temperature over time. You can also download the temperature to a Serial Printer or a PC for further analysis (Models OS533E, OS534E, OS523E, OS524E).

6.After all data is taken, press the LOCK key to unlock the trigger.

2.3Real Time Mode (Active Operation)

Definition: Real Time Mode is the active operational mode of the thermometer. In this mode, the thermometer constantly measures and displays temperature.

FUNC

Figure 2-12. General Operational Block Diagram

2-8

Page 23
Image 23
Omega Engineering OS533E manual Real Time Mode Active Operation, General Operational Block Diagram