Troubleshooting Guide 6

Problem

Solution

The temperature reading is erratic. The thermometer has just been moved from one extreme temperature to room temperature [0°C or 50°C (32°F or 122°F)] or vice versa.

1.The thermometer has to stabilize before taking temperature measurements. It takes up to 30 minutes for the thermometer to stabilize.

The temperature reading is erratic. The thermometer has just been moved from room temperature (ambient temperature) to a temperature 10°C colder or warmer.

1.The thermometer has to stabilize before taking temperature measurements. It takes up to 20 minutes for the thermometer to stabilize.

No Laser Beam

1.

Make sure the trigger is pulled

 

 

and the laser power switch is

 

 

turned on. (The red power LED

 

 

should be lit).

 

 

 

The Laser "line of sight"

1.

The line of sight and the center

does not coincide

 

of the target are offset by design.

with the center of the

 

(refer to Figure 3-4 and the

target.

 

explanation above it for how

 

 

to compensate for this).

6-3

Page 58
Image 58
Omega Engineering OS533E manual Problem