CHANGING THE AIR FILTER

Change the air filter every 30 days even if the air filter does not appear dirty. If the air filter appears dirty, or if water or medication has spilled on the air filter replace with a new air filter immediately.

1.Pull the air filter cover to remove it from the front side of the compressor.

2. Remove the dirty filter with a small pointed object.

CAUTION

Do not attempt to wash or clean the air filter. Damp

air filters can cause blockages. Do not substitute cotton or any other material for the air filter.

CAUTION

Wash the air filter cover regularly to prevent any blockage in the cover. Do not boil. Make sure the cover is dry before inserting the new air filter.

CAUTION

Do not operate without air filter.

3. Insert a new air filter into the air filter cover.

CAUTION

Before inserting the new air filter make sure the air filter is clean and free of dust. Use only the Omron air filter designed for this device.

4. Put the air filter cover back on the compressor.

19

Page 19
Image 19
Omron Healthcare NE-C25 instruction manual Changing the AIR Filter, Insert a new air filter into the air filter cover