GUIDANCE AND MANUFACTURER’S DECLARATION — ELECTROMAGNETIC IMMUNITY:The goLITE BLU is intended for use in the electromagnetic environment specified below.The customer or the user of the goLITE BLU should make sure it is used in such an environment.

Immunity Test

IEC 60601 Test

Compliance

Electromagnetic

 

Level

Level

Environment - Guidance

 

 

 

 

Electrostatic

±(2,4,6) kV contact

±(2,4,6) kV

Floors should be wood,

Discharge (ESD)

±(2,4,8) kV air

contact

concrete or ceramic tile. If floors

IEC 61000-4-2

 

±(2,4,8) kV air

are covered with synthetic

 

 

 

material, the relative humidity

 

 

 

should be at least 30%.

 

 

 

 

Electrical Fast

±2 kV for power

±2 kV for power

Mains power quality should be

Transient/Burst

supply lines

supply lines

that of a typical commercial or

IEC 61000-4-4

±1 kV for input-

±1 kV for input/

hospital environment.

 

output lines

output lines

 

 

 

 

 

Surge

±1 kV differential

±1 kV

Mains power quality should be

IEC 61000-4-5

mode

differential

that of a typical commercial or

 

±2 kV common

mode

hospital environment.

 

mode

±2 kV for

 

 

 

common mode

 

 

 

 

 

Voltage

<5% UT

<5% UT

Mains power quality should

dips, short

(>95% dip in UT)

(>95% dip in

be that of a typical commercial

interruptions

for 0.5 cycle

UT) for 0.5 cycle

or hospital environment. If the

and voltage

40% UT

40% UT

user of the goLITE BLU requires

variations on

(60% dip in UT)

(60% dip in UT)

continued operation during

power supply

for 5 cycles

for 5 cycles

power mains interruptions,

input lines

70% UT

70% UT

it is recommended that the

IEC 61000-4-11

(30% dip in UT)

(30% dip in UT)

goLITE BLU be powered from an

 

for 25 cycles

for 25 cycles

uninterruptible power supply or

 

<5% UT

<5% UT

a battery.

 

(>95% dip in UT)

(>95% dip in

 

 

for 5 sec

UT) for 5 sec

 

Power

3 A/m

3 A/m

Power frequency magnetic

frequency

 

 

fields should be at levels

(50/60 Hz)

 

 

characteristic of a typical

magnetic field

 

 

location in a typical commercial

IEC 61000-4-8

 

 

or hospital environment.

 

 

 

 

NOTE: UT is the a.c. mains voltage prior to application of the test level.