![](/images/new-backgrounds/1017325/1732555x1.webp)
Follow your local protocol with regard to prompt data transfer for medical review after using the HeartStart.* Details about data transfer and timing are provided in Event Review documentation.
The information automatically stored by the HeartStart includes a summary of
•ECG recordings (a maximum of 15 minutes following pads application†)
•the HeartStart’s status (entire incident)
•the HeartStart’s rhythm analysis decisions (entire incident)
•the elapsed time associated with stored events (entire incident)
*The HeartStart automatically stores information about its last clinical use in its internal memory for at least 30 days, so the data can be downloaded to a computer running appropriate Event Review software. (If the battery is removed during this period, the defibrillator retains the files. When the battery is reinstalled, the
†If ECG recordings from a previous use have not been erased, the maximum time for new ECG recordings may be less.