Philips M5066A owner manual Defibrillation Waveform, Category Waveform parameters, Delivered

Models: M5066A

1 70
Download 70 pages 12.08 Kb
Page 45
Image 45

Philips Medical Systems

category

waveform parameters

E-3

Defibrillation Waveform

specifications

Biphasic truncated exponential. Waveform parameters are automatically adjusted as a function of patient defibrillation impedance. In the diagram at left, D is the duration of phase 1 and E is the duration of phase 2 of the waveform, F is the interphase delay (500 µs), and Ip is the peak current.

The HeartStart delivers shocks to load impedances from 25 to 180 ohms. The duration of each phase of the waveform is dynamically adjusted based on delivered charge, in order to compensate for patient impedance variations, as shown below:

 

 

adult defibrillation

 

 

 

load

phase 1

phase 2

peak

delivered

 

resistance (Ω)

duration (ms)

duration (ms)

current (A)

energy (J)

 

25

2.8

2.8

65

128

 

50

4.5

4.5

40

150

 

75

6.25

5.0

30

155

 

100

8.0

5.3

24

157

 

125

9.65

6.4

21

159

 

150

11.5

7.7

18

160

 

175

12.0

8.0

16

158

 

 

pediatric defibrillation

 

 

E

(using M5072A infant/child reduced-energy defibrillator pads)

 

load

phase 1

phase 2

peak

delivered

 

resistance (Ω)

duration (ms)

duration (ms)

current (A)

energy (J)

 

25

4.1

2.8

28

35

 

50

5.1

3.4

20

46

 

75

6.2

4.1

15

52

 

100

7.2

4.8

12

54

 

125

8.3

5.5

10

56

 

150

9.0

6.0

9

57

 

175

9.0

6.0

8

55

 

Page 45
Image 45