Important warnings and reminders

Do not allow the pads to contact other electrodes or metal parts that are in contact with the patient.

Before delivering a shock, it is important to disconnect the patient from other medical electrical equipment, such as blood-flow meters, that may not incorporate defibrillation protections. In addition, make sure the pads are not in contact with metal objects such as a bedframe or stretcher.

Check supplies, accessories, packaging, and spares for damage and expiration dating.

Environmental considerations

The defibrillator contains electronic components. Dispose of it at an appropriate recycling facility.

The battery cells contain chemicals. Recycle the battery at an appropriate recycling facility.

The used pads may be contaminated. Cut them off and dispose of them properly. Recycle the remaining cartridge components at an appropriate recycling facility.

Shock cycle timing

The HeartStart Quick Shock feature allows it to deliver a shock within 8 seconds, typical, following the prompt ending a CPR Interval. From shock to shock, the HeartStart takes <20 seconds, typical, including analysis. After 15 shocks, the HeartStart takes <30 seconds from analyzing to ready-to-shock. After 200 shocks, the HeartStart takes <40 seconds from initial power-on to ready-to-shock.