Patient treatment protocol options

parameter

“call EMS” voice reminder timing

shock series

shock series interval (minutes)

settings

default

default description

• At power on (when

At the start of

Provides a voice reminder to make

the user turns on

the first

sure emergency medical services have

the HeartStart)

patient care

been called, at the start of the first

• At power on and at

pause

patient care pause.

 

 

the start of the first

 

 

patient care pause

 

 

• At the start of the

 

 

first patient care

 

 

pause

 

 

• No reminder

 

 

1, 2, 3, 4

1

The automatic protocol pause for CPR

 

 

is activated each time a shock is

 

 

delivered.*

 

 

During the protocol pause, the

 

 

HeartStart does not perform rhythm

 

 

analysis.

 

 

The length of the protocol pause after

 

 

a shock series is completed is

 

 

determined by the protocol pause

 

 

timer setting.

1.0, 2.0,

1.0

A delivered shock must occur within 1

(infinity)

 

minute of the previous shock to be

 

 

counted as part of the current shock

 

 

series.

 

 

NOTE: This parameter is only applicable

 

 

when the shock series is not configured to

 

 

the default 1 shock.