Philips M5066A NSA pause type Settings, Standard NSA, During a Smart NSA pause, Pause HeartStart

Models: M5066A

1 70
Download 70 pages 12.08 Kb
Page 53
Image 53

Philips Medical Systems

parameter

protocol pause timer (minutes)

NSA pause type

 

 

F-3

 

settings

default

default description

 

0.5, 1.0, 1.5,

2.0

A 2-minute protocol pause for CPR

 

2.0, 2.5, 3.0

 

automatically starts after voice

 

 

 

instruction is given when a shock

 

 

 

series is completed. After the protocol

 

 

 

pause, the defibrillator returns to

 

 

 

rhythm analysis.

 

 

 

If the user presses the i-button for

 

 

 

optional CPR coaching, the HeartStart

 

 

 

provides coaching for 5 cycles of CPR,

 

 

 

starting and ending with compressions,

 

 

 

when the CPR Coaching parameters

 

 

 

are also set to their default values. The

 

 

 

number of CPR cycles varies for other

 

 

 

protocol pause timer and CPR

 

 

 

Coaching parameter settings.

 

 

 

Note: Because the protocol pause ends

 

 

 

upon completion of a CPR cycle in order

 

 

 

to maximize the benefits of CPR, the

 

 

 

actual duration of the pause may differ

 

 

 

slightly from the timer setting.

 

• Standard NSA

SMART NSA

During a SMART NSA pause, the

 

pause: HeartStart

pause

defibrillator conducts background

 

does not perform

 

monitoring. If a potentially shockable

 

rhythm analysis

 

rhythm is detected in a motionless

 

during the NSA

 

patient, the defibrillator terminates the

 

pause.

 

SMART NSA pause and resumes

 

• SMART NSA pause:

 

rhythm analysis.

 

 

 

 

HeartStart con-

 

NOTE: If the HeartStart detects CPR in

 

ducts background

 

progress or if the responder has pressed

F

monitoring during

 

the i-button for CPR Coaching, the SMART

the SMART NSA

 

NSA pause will be converted to a

 

pause. If a poten-

 

standard NSA pause. During the standard

 

tially shockable

 

NSA pause, the defibrillator does not

 

rhythm is detected,

 

perform rhythm analysis.

 

HeartStart termi-

 

 

 

nates the SMART

 

 

 

NSA pause and

 

 

 

resumes rhythm

 

 

 

analysis.

 

 

 

Page 53