4.7 Storing Piece Weight into Memory

There are 10 memory locations (0 to 9 numeric keys) available to store APW values.

After determining the piece weight (refer to Sec. 4.5 or 4.6), long-press APW for approx. 2 seconds. “StorE” will flash on the display.

Press any one of the numeric keys to store the piece weight data in this selected location.

Weight

g

 

 

Piece Weight g/pc

 

Count

pcs

 

0.0

 

 

 

 

0.

 

 

40.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.0

 

 

StorE

 

 

0.

 

 

 

 

 

 

 

 

 

0.0

 

40.0

 

0.

 

 

 

 

 

 

 

 

NOTE: Stored APW values are retained when the unit is turned off.

4.8 Recalling Piece Weight from Memory

Press the numeric key (ex. location 5) with the stored piece weight data (ex. 40 g/pc), then press APW two times. The stored data will be uploaded and displayed.

Weight

g

 

 

Piece Weight g/pc

 

Count

pcs

 

0.0

 

 

 

 

0.

 

 

0.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.0

 

5.

 

0.

 

 

 

 

 

 

 

 

Memory Location

0.0

40.000

0.

4.9 Accumulation: Weight and Quantity

Accumulation is indicated by the

Place the item to be weighed/counted on the pan (ex. 300g with APW of 60g). Press M+. “Add”, then the accumulation event is displayed momentarily before reverting to nor- mal weighing mode.

NOTE: Weight display must return to zero before the next accumulation can register.

To recall Total Accumulation data, press M+ with no weight on the pan. The data is displayed momentarily before reverting to normal weighing mode.

To clear the Total Accumulation data, press C while the data is displayed.

annunciator.

 

Weight

g

 

 

 

 

Piece Weight g/pc

Count

pcs

 

 

300.0

 

 

 

 

 

 

5.

 

 

 

 

 

 

60.0000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

------

 

Add

------

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

300.0

 

=01=

 

5.

 

 

 

 

 

 

 

 

 

Total Accumulated

 

 

 

 

Total Accumulation

Total Accumulated

 

Weight

 

 

 

 

Times

 

Count

 

 

 

 

 

 

 

 

 

 

 

 

300.0

60.000

 

5.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weight

g

 

 

 

 

Piece Weight g/pc

Count

pcs

 

1500.0

 

 

 

 

 

25.

 

 

 

=05=

 

 

 

 

 

 

 

 

 

 

Total Accumulated

 

 

 

 

Total Accumulation

Total Accumulated

 

Weight

 

 

 

 

Times

 

Count

 

 

 

 

 

 

 

 

 

1500.0

 

 

60.000

 

25.

 

 

 

 

 

 

 

 

 

 

NOTE: Accumulation data is not retained when the unit is turned off.

3 EC-textEN.pmd

9

8/9/2005, 1:41 PM

Page 13
Image 13
Recoton/Advent instruction manual Storing Piece Weight into Memory, Recalling Piece Weight from Memory, EC Series EN-9