| EC Series | |
|
|
|
5 SCALE SETTINGS
;
;In Setup mode:
o Press Sample to step through available settings
oPress Tare to accept the displayed setting and proceed to the next Setup parameter
;Press C at any time to exit from the Setup mode.
;
The following parameters are available:
5.1 Auto-Off
Sets the period of inactivity before the scale automatically turns off. The following settings are available: 0 = Off, 2, 5, 8 (minutes)
Weight g
0.
Default Setting
Piece Weight g/pc
A.Off
Count pcs
Internal Count (disregard)
5.2 Zero Tracking Range
Sets the range in which the zero reading is maintained.
The following settings are available: 0 = Off, 1 = 0.5d, 2 = 1d, 3 = 2d, 4 = 3d
(d = scale division)
Weight g
2.
Default Setting
Piece Weight g/pc
trAC
Zero Tracking
Range
Count pcs
Internal Count (disregard)
5.3 Zero Display Range
Sets the range in which the Zero indication turns on.
The following settings are available: 0 = Off, 1 = 0.5d, 2 = 1d, 3 = 2d, 4 = 3d
(d = scale division)
Weight g
3.
Default Setting
Piece Weight g/pc
2Ero
Zero Display Range
Count pcs
Internal Count (disregard)
5.4 Filtering
Sets the level in which the Stable indication turns on; the higher the setting, the faster stabilization time. The following settings are available: 0, 1, 2, 3, 4, 5 (level)
Weight g
1.
Default Setting
Piece Weight g/pc
Filt
Filtering Level
Count pcs
Internal Count (disregard)
3 | 10 | 8/9/2005, 1:41 PM |