Technical specifications
Maximum heater plate temperature | 85°C |
Temperature | 93°C |
Maximum gas temperature | 41°C |
Dimensions (H x W x D) | Docking station and water chamber: 130 mm x 170 mm x 212 mm |
Weight | Docking station and unfilled water chamber: 786 g |
Water capacity | To maximum fill line: 390 mL |
Power supply | |
| Instantaneous power consumption: <400W |
| Average power: <115W |
Docking station | Flame retardant engineering thermoplastic, plated cast aluminium |
Water chamber | Injection molded plastic, stainless steel and silicon seal |
Operating temperature | +5°C to +35°C |
Operating humidity | |
Operating altitude | Please refer to your CPAP, bilevel or |
| user guide |
Storage and transport temperature | |
Storage and transport humidity | |
IEC | Class II (double insulation), Type BF Applied Part |
Electromagnetic compatibility | Please refer to your flow generator user's manual for details |
Humidifier performance
The following settings have been tested at 22°C, 40% RH ambient conditions:
CPAP mask pressure |
| RH output % | Nominal system output AH*, BTPS† | |
cm H2O | Setting 3 | Setting 6 | Setting 3 | Setting 6 |
4 | >95 | >95 | 16 | 20 |
|
|
|
|
|
10 | >85 | >95 | 14 | 18 |
|
|
|
|
|
20 | >75 | >85 | 13 | 16 |
|
|
|
|
|
*AH- Absolute Humidity in mg/L.
†BTPS- Body Temperature Pressure Saturated.
Note: The manufacturer reserves the right to change these specifications without notice.
10