Technical specifications

Maximum heater plate temperature

85°C

Temperature cut-out

93°C

Maximum gas temperature

41°C

Dimensions (H x W x D)

Docking station and water chamber: 130 mm x 170 mm x 212 mm

Weight

Docking station and unfilled water chamber: 786 g

Water capacity

To maximum fill line: 390 mL

Power supply

100-240V 50/60Hz, 0.7-1.6A; 110V 400Hz, 0.8A

 

Instantaneous power consumption: <400W

 

Average power: <115W

Docking station

Flame retardant engineering thermoplastic, plated cast aluminium

Water chamber

Injection molded plastic, stainless steel and silicon seal

Operating temperature

+5°C to +35°C

Operating humidity

10–95% non-condensing

Operating altitude

Please refer to your CPAP, bilevel or non-invasive ventilation device

 

user guide

Storage and transport temperature

-20°C to +60°C

Storage and transport humidity

10–95% non-condensing

IEC 60601-1 classification

Class II (double insulation), Type BF Applied Part

Electromagnetic compatibility

Please refer to your flow generator user's manual for details

Humidifier performance

The following settings have been tested at 22°C, 40% RH ambient conditions:

CPAP mask pressure

 

RH output %

Nominal system output AH*, BTPS

cm H2O

Setting 3

Setting 6

Setting 3

Setting 6

4

>95

>95

16

20

 

 

 

 

 

10

>85

>95

14

18

 

 

 

 

 

20

>75

>85

13

16

 

 

 

 

 

*AH- Absolute Humidity in mg/L.

BTPS- Body Temperature Pressure Saturated.

Note: The manufacturer reserves the right to change these specifications without notice.