Cleaning your H4i

Cleaning your H4i

Water chamber

Daily

12

34

Wash the water chamber, spill guard, tub plate and tub plate seal in warm water, using a mild detergent. Use the yellow cleaning plug to assist in the cleaning process (see steps 1 to 4). Rinse thoroughly in clean water and allow them to dry away from direct sunlight.

Weekly

The disassembled cleanable water chamber, spill guard, tub plate and tub plate seal may be washed in a dishwasher on the delicate or glassware cycle (top shelf only).

Note: Not applicable to the H4i Water Chamber.

Inspect the water chamber for wear and deterioration. Replace the water chamber if any component has cracked, become cloudy or pitted. Replace immediately if there is any leaking from the water chamber.

White powder or deposits in the water chamber can be cleaned using a solution of one part vinegar to 10 parts water, as follows:

1Fill the cleaning plug to the rim with vinegar and pour into empty water chamber.

2Add clean water to the water chamber until the solution reaches the maximum water level mark.

3Let the solution soak for 10 minutes.

4Discard the solution and rinse the water chamber with clean water.