Problem | Possible Cause | Solution |
|
|
|
No/low air flow | The air tubing is not | Connect the air tubing as described in |
| connected properly. | “Setting Up your HumidAire 3i For Use” |
|
| on page 5. |
| The air tubing is blocked, | Unblock or free the air tubing. Check the |
| pinched or punctured. | air tubing for punctures. |
| The flow generator is not | See the Troubleshooting section in the |
| working correctly. | flow generator manual. |
| The water chamber is | Make sure that the water level is not |
| overfilled. | above the maximum water level mark. |
| The cover is not latched | Make sure the cover is latched onto the |
| correctly. | docking station correctly. |
| The cover seal is not fitted | Make sure the cover seal is facing the |
| correctly. | right way up and fitted securely. |
| Condensation is forming in | • Turn the control dial down. |
| the air tube. | • Insulate the air tubing. |
|
| |
|
|
|
Water splashing on | The water chamber is | Empty some water from the water |
your face | overfilled. | chamber. |
| Condensation is forming in | • Turn the control dial down. |
| the air tube and mask. | • Insulate the air tubing. |
|
| |
|
|
|
Air does not feel | The control dial is set too | Turn the control dial up. |
moist | low. |
|
| HumidAire 3i heating | Contact your equipment supplier. |
| element is not working. |
|
|
|
|
Air too moist | Control dial is set too high. | Turn the control dial down. |
| High relative humidity in | Adjust your humidity setting. |
| your room. |
|
White powder or | The water leaves scaly |
deposits in the water | mineral deposits. |
chamber |
|
•Use distilled water.
•To remove these mineral deposits, see “Cleaning your HumidAire 3i” on page 11.
14