ResMed s8 user manual Troubleshooting, Problem, Possible Cause, Solution

Models: s8

1 47
Download 47 pages 54.64 Kb
Page 34
Image 34
TROUBLESHOOTING

TROUBLESHOOTING

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem

Possible Cause

Solution

 

 

 

No display.

Power not connected.

Ensure the appropriate power

 

 

cord is connected to your S8 Elite

 

 

and the power switch (if

 

 

available) is on.

 

 

 

Insufficient air

Ramp time is in use.

Wait for air pressure to build up

delivered from

 

or change ramp time.

S8 Elite.

Air filter is dirty.

Replace air filter.

 

 

Air tubing is kinked or

Straighten or replace tubing.

 

punctured.

 

 

Air tubing not connected

Check air tubing.

 

properly.

 

 

Mask and headgear not

Adjust position of mask and

 

positioned correctly.

headgear.

 

Cap(s) missing from access

Replace cap(s).

 

port(s) on mask.

 

 

Pressure required for treatment

See your clinician to adjust the

 

may have changed.

pressure.

 

 

 

S8 Elite does not start

Power cord not connected

Connect power cord firmly at

when you breathe into

properly.

both ends.

the mask.

 

 

ENGLISH

Power outlet may be faulty.

SmartStart not on.

Leak Alert has been enabled; SmartStart has automatically been disabled.

Use of a Mirage or Ultra Mirage Full Face Mask.

Breath is not deep enough to trigger SmartStart.

Try another power outlet.

Enable SmartStart.

Disable Leak Alert to enable SmartStart.

SmartStart does not work with a full face mask as the anti- asphyxia valve will not allow sufficiently high pressure on exhalation.

Take a deep breath in and out through the mask.

TROUBLESHOOTING 33

Page 34
Image 34
ResMed s8 user manual Troubleshooting, Problem, Possible Cause, Solution