Problem | Possible Cause | Solution |
|
|
|
| There is excessive leak. | Adjust position of mask and |
|
| headgear. |
|
| Caps may be missing from ports |
|
| on mask. Replace them. |
|
| Air tubing not connected |
|
| properly. Connect firmly at both |
|
| ends. |
|
| Air tubing kinked or punctured. |
|
| Straighten or replace. |
|
|
|
S8 Elite unit does not | SmartStart/Stop is disabled. | Enable SmartStart/Stop. |
stop when you remove |
|
|
your mask. |
|
|
|
|
|
SmartStart is enabled | Incompatible humidifier or mask | Use only equipment as |
but the flow generator | system being used. | recommended and supplied by |
does not stop |
| ResMed. |
automatically when |
|
|
you remove your |
|
|
mask. |
|
|
Display error message: | The air tubing is loose or |
Check tube!! | blocked. |
Key if done |
|
Check that the air tubing is connected securely to your mask and the air outlet on the front of the S8 Elite.
When you have checked the air tubing, press the Start/Stop key to restart the device.
If this does not clear the message, disconnect the power cord and then reconnect it to restart the device.
Display error message: | Component failure. | Return your S8 Elite for servicing. |
SYSTEM ERROR
Call service!
Display error message: | You have experienced | |
High | leak in | excessively high leak levels |
last | session | during the night. |
Check that your air tubing is connected properly and that your mask does not leak excessively. Use the
If this message appears again, contact your clinician.
34