ResMed s8 user manual System Specifications, Power outlet for HumidAire

Models: s8

1 47
Download 47 pages 54.64 Kb
Page 38
Image 38
SYSTEM SPECIFICATIONS

SYSTEM SPECIFICATIONS

S8 ELITE (NO ACCESSORIES)

Performance: Operating pressure range: 4 to 20 cm H2O

Dimensions (H x W x D): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)

Weight: 2.9 lb (1.3 kg)

Power Supply

Input range for S8 ELITE with HUMIDAIRE 3i: 100–240 V, 50/60 Hz; 110 V, 400 Hz; 2.5 A <140 VA (110 W) (maximum power consumption). Instantaneous peak power consumption <340 VA.

Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed DC-12 converter for DC input.

Actual power consumption will vary, depending on factors such as the use of accessories, height above sea level, and ambient temperature. Power consumption values for typical treatment conditions (using a breathing machine set to 0.5 L and

15breaths/min with 6’6” (2 m) air tubing, and an ULTRA MIRAGE mask) are given in the following table

Device

Pressure

Power

Power

(cm H2O)

(VA)

(W)

 

S8 Elite

5

18

7.6

 

 

 

 

 

10

23

11.6

 

 

 

 

 

15

30

15.9

 

 

 

 

 

20

39

20.7

 

 

 

 

Power outlet for HumidAire 3i

Outlet range: 100–240V, 50/60Hz; 110V, 400Hz; 2.0A

Air Filter: Two-layered, powder-bonded, polyester non-woven fiber

Air Tubing: Flexible plastic, 6’6” (2 m)

Air Outlet: The 22 mm conical air outlet complies with EN 1281-1

IEC 60601-1 Classifications: Class II (double insulation), Type CF

Noise level: Radiated sound pressure is measured at <30 dBA at 10 cm H2O and 1 m (3’3.37”) according to ISO17510

ACCESSORIES

DATA CARD

Dimensions (L x W x D): 2.56” x 0.87” x 0.03” (65.04 mm x 22.15 mm x 0.85 mm)

ENGLISH

SYSTEM SPECIFICATIONS