Clinical Use Only

English

Disinfecting/Sterilizing the Mask or Nasal Pillows System between Patients

This Disinfection/Sterilization Guide is intended for multipatient use of the mask/nasal pillows* system in a sleep lab, clinic or hospital environment. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions.

The following are the recommended and validated procedures for cleaning, disinfection and sterilization of the mask. We advise that each healthcare facility consult its own procedures before carrying out the following instructions.

For details on the mask refer to the User Guide. Parts that cannot be disinfected/sterilized are listed in the table on page 5.

WARNINGS AND CAUTIONS

ResMed cannot give any assurance that deviations from the procedures listed below, and their effect on the performance of the product, will be acceptable.

The mask components should not be subjected to autoclave or ethylene-oxide gas sterilization.

When using detergents, disinfectants or sterilization agents, always follow the manufacturer’s instructions.

Notes:

The headgear does not require disinfection or sterilization. However, it is necessary to thoroughly wash the headgear before using between patients. The cleaning procedure is detailed in the User Guide.

The mask must be disassembled and cleaned prior to disinfection or sterilization. The mask must also be inspected after disinfection/sterilization. See the section "Inspection after Disinfection/Sterilization" for inspection details.

* References to mask/s in this guide refer to ResMed masks and nasal pillows system.

1

Page 2
Image 2
ResMed Mirage Vista, Ultra Mirage II, Mirage Kidsta, Mirage Swift II, Mirage Activa, Mirage Series 2 manual English