NOTE: Monitor subsequent films on the same steps selected for normal contrast index.
•Date - Month - Day – Year
•Processor Number - Processor identification
•Emulsion Number - Film batch identification
•Developer Type - Developer vendor identification
•Fixer Type - Fixer vendor identification
•Film Type - Film vendor identification
•Exposure Color - Exposure light (blue or green)
•Developer Replenisher Rate - The rate of developer replenishment
•Fixer Replenisher Rate - The rate of fixer replenishment
•Processing Time - Film process time, input-to- output
A box of film should be set aside from regular stock for exclusive sensitometer use. New film stock will require
Daily Plotting of Data on Process Control Record
Plot the Speed Index, Contrast Index, Base+Fog, and Developer Temperature results on the process control record each time a control film is developed. Record data immediately so that it is not lost or changed.
8