activated. When the green color is selected, switches A,B,C,D bracketed as green are activated.
Exposure Setting | A | B | C | D | |
1 | MIN | OFF | OFF | OFF | OFF |
2 |
| ON | OFF | OFF | OFF |
3 |
| OFF | ON | OFF | OFF |
4 |
| OFF | OFF | ON | OFF |
5 |
| ON | ON | ON | OFF |
6 |
| OFF | OFF | OFF | ON |
7 | MAX | OFF | OFF | ON | ON |
Adjusting Exposure Setting (394)
The sensitometer has the capability of seven different exposure times for both single and dual. The factory preset exposure setting is “3” for both single and dual. If different exposure times are required, refer to the chart
The dipswitch used to change the exposure time is located on the bottom of the unit. When single emulsion is selected, switches A,B,C,D bracketed as “SINGLE” are activated. When the dual emulsion is selected, switches A,B,C,D bracketed as “DUAL” are activated.
Exposure Setting | A | B | C | D | |
1 | MIN | OFF | OFF | OFF | OFF |
2 |
| ON | OFF | OFF | OFF |
3 |
| OFF | ON | OFF | OFF |
4 |
| OFF | OFF | ON | OFF |
5 |
| ON | ON | ON | OFF |
6 |
| OFF | OFF | OFF | ON |
7 | MAX | OFF | OFF | ON | ON |
13