6.Release the expose button and remove the film immediately.
7.Develop the film in the processor to be monitored (refer to Section 2).
8.Record data on film immediately after development (refer to Section 2).
IMPORTANT NOTES:
•Film must be inserted all the way to the back of the unit in order to be exposed properly.
•The sensitometer is calibrated to expose screen- type films normally used for general radiography to an approximate density of 1.OD+ Base+Fog on Step No. 11.
•If the
-The exposure must have a gamma of .7 or greater on steps 7 through 15. (increasing density increments of at least .11D).
-The film must have at least 1.25 inch of clear leader at both ends of the exposure. The use of 8" x 10" or 18cm x 24cm film is recommended.