BEFORE COMMENCING OPERATION

BEFORE COMMENCING OPERATION

Preparation of monitor bottle

Prior to the operation of the unit, fill the monitor bottles (upper and lower) with 10% glycerol or water using the following procedure:

1.Remove the top and bottom drawer (or shelf).

2.Remove the monitor bottle cover as shown in the figure 1.

3.Take out the sensor from the monitor bottle.

4.Remove the fixing screw and take out the monitor bottle.

5.Remove the cap by loosing it as shown in the figure 2.

6.Pour 200ml of 10% glycerol or water into the monitor bottle.

7.Cap the monitor bottle and replace the bottle to its original position. Then set the sensor to the bottle and replace the cover on the bottle.

Recorder sensor

 

 

(lower bottle of MBR-304G

 

Temp. sensor

and MBR-304GR)

 

Fixing screw

Cover

Fig. 1

Cap

 

 

 

Loosen

Fig. 2

13

Page 14
Image 14