Sanyo MBR-304DR instruction manual Remote alarm terminal, Before Commencing Operation

Models: MBR-304DR

1 26
Download 26 pages 8.47 Kb
Page 15
Image 15
Remote alarm terminal

BEFORE COMMENCING OPERATION

Remote alarm terminal

The terminal for remote alarm is installed inside the rear cover of the unit. The alarm is outputted from this terminal.

The contact rating is DC 30V, 2 A.

Contact output: At normal condition “Open” At abnormal condition “Close”

Note:

The contact is “Close” because it is regarded as abnormal when the power supply cord is disconnected from the outlet or the power switch is turned off, or in the event of power failure.

The remote alarm is in conjunction with the alarm buzzer, the alarm can be silenced by pressing the alarm buzzer stop key (BUZZER) except for in the case of power failure

Terminal